Products
That's us
Products
Downloads
Dates & Addresses

 

Photometer 680

Made in Germany

TECHNICAL SPECIFICATIONS

Type

Single-beam filter photometer

Light Source

Tungsten spotlight

Wavelength

340 nm - 900 nm

Wavelength Selection

Interference filters in discrete holders

Photometric Range

0 - 2.5 A

Cuvette System

Normal standard cuvette (macro or semi-micro, disposable or special optical glass)
optional: round cuvette (16 mm diameter)

Operator Interface

Touchscreen for direct functions and alphanumerical inputs

Data Presentation

Graphic display: white characters or symbols, blue background, lighted, resolution 240 * 128 dots

Languages

English and French/German/Indonesian/Russian/Spanish

Memory

  • General operating software can be updated by PC
  • Reagent-open system with capacity for up to 231 programmable methods
  • Data import by touchscreen or PC
  • easy to use by pre-programmed parameters
  • Up to 50 nonlinear calibration curves with max 20 sets of points can be stored

Signal Port

Serial port for connection to an external printer and/or PC

Data Logging

Up to 1000 results can be saved in the memory automatically

Measurement Procedures

  • Absorbance
  • End point with factor, standard or multiple standards, with or without reagent blank and/or sample blank
  • Turbidimetry with optional timer function
  • Single, double and triple determinations
  • Curve fitting for nonlinear standard curves

Quality Control

Up to 50 methods can be controlled with two control serums, Levey Jennings plot

Mains Supply

Input: 100 VAC up to 240 VAC at 50/60 Hz, output: 12 VDC / 1 ADC

Weight

1.1 kg

ROBERT RIELE GmbH & Co KG • Kurfuerstenstrasse 75-79 • 13467 Berlin • Germany   

Imprint/Impressum (§ 6 TDG)  Disclaimer  Updated: Thursday, 27-Oct-2011 16:58:40 CEST