TECHNICAL SPECIFICATIONS

      

Type

Single-beam filter photometer

Light Source

Halogen lamp - 12 V, 20 W with protection for lifetime

Wavelength

340 nm - 800 nm

Wavelength Selection

Automatic via 9-position filter wheel :
6 standard interference filters: 340 nm, 405 nm, 492 nm, 546 nm, 578 nm, and 623 nm;
3 positions for optional filters of choice

Photometric Range

0 - 2.5 A

Cuvette System

Normal standard cuvettes (macro or semi-micro, disposable or special optical glass)

Temperature Control

Measuring cell as well as dry incubation block with 8 positions temperature controlled by electrical thermostat at 37 °C

Operator Interface

Touchscreen for direct functions and alphanumerical inputs

Data Presentation

Graphic display: White characters or symbols, blue background, lighted, resolution 240 * 128 dots

Integrated Printer

Dot-matrix printer, 24 characters per line

Languages

English and German / Indonesian / Russian / Spanish

Memory

  • General operating software can be updated by PC
  • Reagent-open system with capacity for up to 231 programmable methods
  • Data import by touchscreen or PC
  • Up to 50 nonlinear calibration curves with max 20 sets of points can be stored

Signal Port

Serial port for connection to an external printer and/or PC

Data Logging

Up to 2970 results of max 99 samples can be managed in the memory

Measurement
Procedures

  • Absorbance, Transmission
  • Endpoints with factor, standard or multiple standards, with or without reagent blank and/or sample blank
  • Bichromatic end point
  • Kinetics with standard, factor or multiple standards, with or without blank
  • Fixed time with standard, factor or multiple standards, with or without reagent blank
  • Turbidimetry with optional timer function
  • Single, double and triple determinations
  • Curve fitting for nonlinear standard curves
  • Free hemoglobin in combination with optional interference filters

Quality Control

Up to 50 methods can be controlled with two control serums, Levey Jennings plot

Measuring Time

  • Kinetic: variable from 3 - 25 deltas, time per delta 4 - 255 s
  • Fixed time: variable from 0 - 65535 s

Delay Time

Programmable from 0 - 65535 s

Mains Supply

Range: 90 VAC up to 264 VAC at 50/60 Hz

Dimensions

Length 33 cm x width 34 cm x height 18 cm

Weight

6 kg

ROBERT RIELE GmbH & Co KG • Kurfuerstenstrasse 75-79 • 13467 Berlin • Germany   